徠卡顯微鏡系統
雷射掃描共軛焦顯微鏡
半導體及工業用顯微鏡
顯微數位影像系統
細胞基因影像分析系統
其它光電微機電系統
其它生醫研究儀器
維修與技術支援服務
聯絡與留言

Surface Plasmon Resonance Instruments

表面等離子體共振光譜儀

 

SR7000DC Dual Channel Spectrometer

  • Simultaneous Two Channel Surface Plasmon Resonance Measurements
  • Reference Channel Minimizes Non-Specific Binding and Bulk Refractive Index Effects
  • High Speed Scanning for Fast Kinetics
  • Sensitivity: 0.35 micro Refractive Index Units rms
  • Low Drift for Long Association/Dissociation Times

Exclusive Features of Reichert SPR Instruments


SR7000 Single Channel SPR Spectrometer

  • Monitors Surface Plasmon Resonance and Critical Angle Responses
  • Response Vs. Time and Reflectivity Data
  • Macromolecular to Small Molecule Binding
  • Broad Refractive Index Range
  • Sensitivity: 0.75 micro Refractive Index Units rms.

  • Label-Free, Real-Time Detection of Molecular Interactions
  • Flexible Open Architecture Systems
  • Easy to Configure, Reconfigure, Upgrade
  • Excellent Baseline Stability
  • Broad Range of Flow Rates
  • Low Maintenance Costs

     

 

System Components
  • Manual Injector
  • Semi-Automatic Injector
  • Autosampler
  • Fluidic Components
Sensor Chips
  • Unmodified Gold
  • Planar – 2 Dimensional
  • High Density – 3 Dimensional
  • Broad Range of Biomolecule Immobilization Strategies
     

Software

  • Data Collection and Analysis Software
     

 

Principles of Surface Plasmon Resonance

SPR is a Quantum Mechanical Phenomenon

Snell’s law describes what happens when light is directed through a high refractive index prism (e.g. Sapphire - refractive index 1.76) to a surface of the prism in contact with a low refractive index medium, for example physiological buffer (See Figure 1). Light rays below the critical angle will exit the prism bending toward the prism surface. Light rays above the critical angle are totally internally reflected back through the prism. Photons that are totally internally reflected create an electric field at the interface. Light is not coming out of the prism but an electric field extends past the reflecting surface. This field oscillates with the usual characteristics of an electromagnetic mode. The electrical component perpendicular to the interface decays exponentially – this is called an evanescent wave. This wave is bound to the surface and is used in SPR to detect changes occurring at the surface.

SPR spectroscopy adds a thin metal layer between the prism surface and the aqueous compartment (See Figure 2). Free electrons in the metal layer can act as a resonator. Energy for the resonance comes from the evanescent wave produced by the totally internally reflected photons.

When certain conditions, determined by the wavelength, illumination angle, and refractive index of the prism, metal and aqueous layers are met, coupling/resonance occurs between the plasma oscillations of the free electrons in the metal and the bound electromagnetic field of the totally internally reflected photons. This coupling is the result of the momentum of the incoming light equaling the momentum of the plasma electromagnetic field. Photons are “absorbed” and converted to surface plasmons. Since the photons are not reflected, a “shadow” occurs in the reflected light.

Reichert Inc.’s SPR instruments are optically configured to illuminate a spot on the gold surface over a range of angles from about 58 to 85 degrees. When the aqueous compartment is e.g. physiological saline solution the “shadow” or SPR minimum occurs at about 66 Degrees. Mass changes at the interface between the gold layer and the aqueous compartment cause changes in the local refractive index near the gold layer. This changes the coupling/resonance angle. For example the if a protein layer was added to the gold / aqueous interface the coupling angle would be about 66.6 Degrees. Actual reflectivity from the SR7000 (Angle vs. reflectivity) before and after immobilizing a protein to the surface is shown in Figure 3.

In the SR7000 and SR7000DC Spectrometers the gold surface is illuminated with a range of angles that encompass these shifts in the coupling angle. This range of angles is continuously monitored with a linear photodiode array detector. Image analysis determines the angle at which the reflectivity minimum occurs. This is continuously sent to the data acquisition computer/software.

 

 

“The more success that quantum theory has, the sillier it looks”.

-Albert Einstein to Heinrich Zangger on Quantum Theory, May 20, 1912

“All these fifty years of conscious brooding have brought me no nearer to the answer to the question, 'What are light quanta?' Nowadays every Tom, Dick and Harry thinks he knows it, but he is mistaken. “

-Albert Einstein, 1954.



Figure 1

At incident angles below the critical angle light is refracted out of the prism, angle angles above the critical angle light is totally internally reflected back through the prism.
 


Figure 2

Adding a thin gold layer creates a “resonator” for the incoming light. This resonance occurs at an angle dependent on the refractive index or mass in contact with the solution side of the gold layer. As more mass is added to the surface, e.g. during an antigen binding to a surface immobilized antibody, the resonance angle “shadow” increases.
 

 


Figure 3

Reflected light image as viewed by the SR7000DC Photodiode Array image detector. The resonance angle increases as protein is added to the surface. This angle change is monitored and recorded in real-time.

Surface Plasmon Resonance Applications/Protocols

SR7000DC Application Examples

Application Note I - Antibody/Antigen Interactions System (PDF 160 kb)
Application Note II - Small molecule Binding Assay (PDF 144 kb)
Application note III - Thermodynamic Investigation of an Enzyme-Inhibitor Pair (PDF 156 kb)
Application Note IV - Small Volume Injections with the Reichert Pump (PDF 140 kb)
Application V - Combined Electrochemistry & SPR to Monitor an Electropolymerization
(PDF 144 kb)

Steps of a Typical SPR Experiment:

  1. Clean Instrument
  2. Immobilize Ligand
  3. Inject Analyte (binding partner)
  4. Regenerate Surface
  5. Analyze Data

Immobilization Strategies (PDF 27.9 kb)

Surface Plasmon Resonance Software

Software

Data Acquisition and Alignment

SPR_V4017 Data Acquisition and Alignment Program

The software collects refractive index response, temperature and timed event marks vs. time data from the SR77000 or SR7000DC. This program is also used to set instrument operating parameters. Files are tab delimited text files and can be opened in standard spreadsheet programs. The data acquisition program displays each channel individually, both channels or the difference between channels. The data acquisition interface is shown in Figure 1A. This program is also used to extract and overlay a series of response vs. time curves – see Figure 1B. Overlaid / normalized Data can then be saved as a tab delimited file in formats compatible with global fitting programs “Clamp” or “Scrubber™” (see Biologic Software and The Center for Biomolecular Interaction Analysis http://www.cores.utah.edu/interaction/). Data is also easily imported into standard analysis programs, e.g. Graphpad Prism™, Origin™ or CurveExpert™. Figures 2 a - c show typical data analysis steps using “Scrubber™“ for the interaction of human serum albumin (HSA) with immobilized anti-HSA IgG.

SPR_V4017 also controls the SR7300 semi-automatic injection valve.

 
Figure 1A: SPR_V4017 data acquisition interface
 
Figure 1B: SPR_4017 “Load and Align” form

 


Data Analysis – Scrubber 2®

Links:

Scrubber 2 Software

Additional information on Biomolecular Interactions

Tutorial on theory of data analysis
 

Scrubber 2 is a commercial program of Biologic Software. The program can read and write the SR7000 and SR7000DC data files. The purpose of the program is to align and clean the response vs. time data and globally fit data to biointeraction models to determine reaction kinetics and/or equilibrium constants. Scrubber critically aligns response curves and renormalizes the response and time scales to zero. It can remove bulk refractive index shifts from the SR7000 single channel SPR Spectrometer. Scrubber allows users to set separate start and stop times for each injection – this is important especially with manual systems where start and stop times may vary slightly. It can also correct for bulk refractive index shifts introduced by use of DMSO to dissolve samples that are not easily dissolved in running buffer alone. It also subtracts reference channel response (SR7000DC) and buffer blanks.

 
Figure 2A: Raw overlaid data
 

Figure 2B: Time zeroed, aligned, referenced and buffer blank subtracted data

     

Figure 2C: Global data fit to 1:1 binding model


Reichert Autosampler

Our software precisely controls the Injection (Association) and Load (Dissociation) times for all sample injections. Injection Volumes, the Syringe speed (you adjust slower for more viscous samples), the needle height and whether or not the sample is picked up using headspace pressure and an air segment can also be controlled by the user.

The software can be programmed to refill the syringe on a syringe pump for continuous overnight sampling. A separate Maintenance menu allows the user to perform needlewashes.

Systems, Methods and Run Tables can be created, saved, and exported and imported from one system or computer to another.
 

美嘉儀器股份有限公司 Major Instruments Co., Ltd.
台北總公司 : TEL : 02-2808-1452  FAX : 02-2808-2354  維修中心 : 02-2808-2353   高雄辦事處 - TEL : (07) 727-6569   FAX : (07) 725-8204

E-mail : major@major.com.tw    http://www.major.com.tw